For this purpose, the authors evaluated HYDRAGEL 18 A1AT Isofocusing kit on the semi-automatic HYDRASYS System (Sebia) for the determination of A1AT phenotypes by isoelectrofocusing on ready-to-use agarose gels with specific immunological detection.
The good within-gel reproducibility of this kit was demonstrated using two normal serum samples (M1 and M1M2 phenotypes) and six pathological serum samples with different phenotypes (MS, SS, SZ, MZ, ZZ). A sensitivity study was undertaken by performing serial dilutions on a serum with a ZZ phenotype containing 0.27 g/L A1AT. The detection limit was 0.050 g/L.
The authors concluded that HYDRAGEL 18 A1AT Isofocusing kit is highly specific, rapid and simple to perform. It improves identification of not only the most common but also various rare A1AT phenotypes. It appears to be suitable for routine analysis and screening applications in a clinical laboratory setting.
Test associated with this resource
Related instrument
Ampla gama de opções com resultados de alta qualidade na eletroforese em gel.
Discover more resources on Alpha-1 Antitrypsin Deficiency
Software abrangente de análise e interpretação de dados para ensaios da Sebia em instrumentos de eletroforese.
Esta secção contém informações destinadas a ampla divulgação e, por conseguinte, pode conter detalhes sobre produtos ou informações que não estejam disponíveis ou não sejam válidas no seu país. Contacte o seu representante local da Sebia. Informações destinadas a profissionais de saúde. Leia atentamente as instruções constantes das bula dos reagentes e dos manuais dos instrumentos.