A method comparison study involving 103 patient samples demonstrated excellent agreement between Sebia's CAPILLARYS 3 OCTA® and the ADAMS™ HA-8180T HPLC system for HbA1c measurement. With a correlation coefficient of 0.991 and minimal bias, the findings support the reliable use of CAPILLARYS 3 OCTA® for routine HbA1c testing and highlight its ability to ensure consistent patient monitoring when laboratories transition between analytical methods.
Evolution makes the difference.
Comprehensive data analysis and interpretation software for Sebia assays on electrophoresis instruments.
This section contains information intended for wide distribution and may therefore contain product details or information that is not available or valid in your country. Please contact your local Sebia representative. Information intended for healthcare professionals. Carefully read the instructions in the reagent package inserts and instrument manuals.